Figure 11-14 DGS/AVG setup page
The
DGS/AVG
setup page allows you to define the exact probe being used for the
inspection, and to set up the DGS/AVG curve to be drawn. There are several
adjustments that can be made from this setup page:
—
DGS/AVG
Activate/Deactivate the DGS/AVG function
—
Probe Type
Selects the type of probe to be used (Straight Beam, Angle Beam, or Dual
Probe).
—
Probe Name
Selects the actual probe to be used.
—
Reflector Type
Defines the type of reflector to be used to acquire a reference indication to
build the DGS/AVG curve. For straight beam and dual probes, the available
reflectors are:
•
Back wall
•
Side-drilled hole (SDH)
•
Flat-bottom hole (FBH)