DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
230
To build a DAC curve
1.
Move the probe to the first echo.
2.
Press
G1 Start
(P5 key) and center gate 1 over the echo.
3.
Press
Add
(P1 key).
4.
Press 2
ND
F, (AUTO XX%) to activate AUTO 80 % for the indication.
You can press
Delete
(P2 key) to delete a point that is incorrectly acquired.
5.
Move the probe to the next echo and center gate 1 over the echo.
You can adjust the RANGE and DELAY while detecting the echo. You can also go
to the Trig menu item and adjust the
Angle
,
Thick
, and
CSC
parameters.
Once a point is captured, the peak amplitude of that point is marked using a “x”
symbol. Figure 11-4 on page 230 shows a partially completed DAC curve.
Figure 11-4 DAC Setup one point
6.
Press 2
ND
F, (AUTO XX%).
7.
Repeat capturing until a curve is built (See Figure 11-5 on page 231.)
8.
Press
Done
(P3 key).
Summary of Contents for EPOCH 650
Page 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Page 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Page 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Page 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Page 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Page 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Page 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Page 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Page 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Page 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Page 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Page 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...