DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
136
3.
While the probe is still on a good part of the standard, press the CAL NULL key
(
).
4.
Scan over the defects and ensure that the signal goes outside the alarm box.
Readjust the
GAIN
as needed.
5.
Press the FULL NEXT key (
) to display the full screen, and then scan over
defects again (see Figure 6-20 on page 136).
Figure 6
‑
20 The full
‑
screen display of the scan
To fine-tune the instrument settings
1.
Depending on your requirements, set the alarm box parameters, horn, or external
horn (louder). For more details about alarms, see “Alarms, Connectivity, and
Memory Specifications” on page 217.
2.
Depending on your requirements, change your real-time readings.
The default real-time readings display the XY signal’s live peak-to-peak
amplitude. For information on how to turn off the real-time readings, see
“Displaying Real-Time Readings” on page 58.
The list of all parameters is shown in Figure 6-21 on page 137.
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...