DMTA-10045-01EN, Rev. E, August 2016
Control Functions
107
The BondMaster 600 is capable of storing and retrieving complete instrument setups.
By default, the data is saved with the date, time, and an instrument-generated file
name. If a PowerLink probe remains connected when the data is saved, the probe part
number and description are also recorded.
After the data is stored, you can give the file a new name, which may contain up to 29
alphanumeric characters. Notes can also be added to the file.
Changes to the file name and notes are made using the front panel of the
BondMaster 600, or alternatively, by using the BondMaster PC software (included
with each instrument).
Whenever a program (a stored data file) is recalled, the currently active instrument
settings are overwritten and cannot be recovered, unless these settings were
previously stored in another program location.
The following memory storage (
MEM
) functions are available:
PREVIEW
You can use
PREVIEW
to view instrument screen images that were captured at
the moment the data was saved. To preview a stored data file, press the MEM
menu key (
), rotate the knob until the desired data file is highlighted, and
then press the A key. The instrument screen image that was stored when the data
file was saved will appear on the BondMaster 600 screen. The following actions
can then be performed:
— Go back (return to the previous menu) by pressing the A key.
— Recall the stored data file by pressing the B key.
— Set the data file as a reference image by pressing the D key.
RECALL
This function resets the BondMaster 600 and loads the instrument settings
associated with the recalled data file.
To recall a stored data file, press the MEM menu key (
), rotate the knob until
the desired data file is highlighted, and then press the B key. The BondMaster 600
recalls the data file with the instrument settings that were stored when the data
file was saved.
NOTE
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...