DMTA-10045-01EN, Rev. E, August 2016
Applications
135
Figure 6
‑
18 The Skin To Core Disbonds (Tapered) application
To calibrate the signals
1.
Press the MAIN menu key (
) once, and then use the knob to set the
START
FREQ
(C key) to
10 kHz
and the
STOP FREQ
(D key) to
40 kHz
.
2.
Place the probe tips on a good part of the standard, press the GAIN key (
)
and then use the knob to adjust the gain until the swept figure is contained
between two divisions inside the alarm box (see Figure 6-19 on page 135).
Figure 6
‑
19 The swept figure between two divisions