General Information

Warmer Operation

When operating as a radiant warmer, select the control mode
by pushing the Manual Mode Button in the middle or the Baby
Mode Button at the top of the controls.

Manual Control Mode

The Manual Mode Indicator lights when the manual mode is
selected. The power setting is indicated by the bar graph LED
in the middle third of the display. The percentage of full power
can adjusted in 5% increments using the temperature/power
buttons located to the left of the mode buttons.

Note: when operating the OmniBed as a warmer in the
manual mode with a probe attached to the infant’s skin, the
infant’s temperature will also be shown in baby temperature
section at the top of the display.

Baby Control Mode

The Baby Mode Indicator lights when baby mode is selected.
Two temperature values are shown in the upper third of the
display: the value shown in larger numerals at the top is the
temperature measured by the probe on the infant’s skin, the
value shown below in smaller numerals is the baby set tem-
perature. The set temperature can be adjusted by using the
temperature/power buttons located to the left of the mode
buttons. Temperature settings from 35° to 37°C can be entered
in 0.1°C increments. Temperature settings from 37° to 37.5°C
can only be entered after pressing the >37°C Button.

Note: when operating the OmniBed as a warmer in the baby
mode, the bar graph in the middle of the display will also
indicate the percentage of power at which the heater is
currently running.

Operating with Two Temperature
Probes

The OmniBed is equipped with two patient temperature probe
jacks for co-bedding of twins, or to monitor a single baby’s
temperature from two anatomical sites. They are identified as
probe 1 and probe 2 on the jack panel behind the bed. Tem-
perature readings for probe jack 1 are displayed in Baby
Temperature at the top of the numeric displays. Temperature
readings from probe jack 2 are displayed in the upper right
corner of the graphic display labeled Probe 2.

The OmniBed will only Baby Control from a probe attached to
jack 1. When both jack 1 and jack 2 are used, the unit will not
operate in the Baby Control mode.

CI.31.01

1

1. Manual Mode Button
2. % Radiant Heater Power

25

%

50

%

75

%

100

%

1

2

1. Baby Mode Button
2. Baby Temperature
3. Baby Set Temperature

25

%

50

%

75

%

100

%

CI.31.012

1

2

3

?

Options
Done/Exit

Probe 2

36.8

Alarm messages displayed here

10:30 am

6/23/00

Name

Birth: 6/21/00

Wed

1. Probe 2 Temperature

1

CI.31.013

Summary of Contents for Giraffe OmniBed

Page 1: ...GiraffeTM OmniBedTM Operator s Manual...

Page 2: ...Center This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Ohmeda Medical and by Ohmeda Medical trained personnel The Product must no...

Page 3: ...r 3 3 Operation as a Warmer 3 4 Attaching the Skin Temperature Probe 3 4 Additional Operating Procedures 3 5 Accessing the Baby 3 5 Raising the Canopy 3 6 Raising and Lowering the Bed 3 6 Tilting the...

Page 4: ...you are specifically instructed to do so as part of the procedure Thoroughly air dry the OmniBed after cleaning it with flammable agents Small amounts of flammable agents such as ether alcohol or sim...

Page 5: ...t The baby can be accessed through portholes or doors When you require even greater access to the baby the canopy can be raised and the com partment walls can be lowered As the canopy raises doors ope...

Page 6: ...was used in the last 2 hours it will ask if you would like to retain the settings you used previously If you are using the OmniBed as an incubator canopy closed you select baby or air mode If you are...

Page 7: ...numerals is the baby set tempera ture which can be adjusted by using the temperature power buttons located at the left of the mode buttons Temperature settings from 35 to 37 C can be entered in 0 1 C...

Page 8: ...located to the left of the mode buttons Temperature settings from 35 to 37 C can be entered in 0 1 C increments Temperature settings from 37 to 37 5 C can only be entered after pressing the 37 C Butt...

Page 9: ...y mode using the same baby set temperature you entered earlier If the canopy is lowered while the OmniBed is being operated in the manual mode as an warmer it will operate in the air mode as an incuba...

Page 10: ...n Done Exit 10 30am 6 23 00 Name Birth 6 21 00 Wed ALARMS HighAirTemp LowAirTemp BabyCold BabyHot AirTemp 38C AirTemp 40C down Done Exit BABYCOLD 1 0CunderBABYSETTEMP Tempprobepoorlyattachedtobaby Reh...

Page 11: ...rended period and heater power percentage if it was operated as a radiant warmer By selecting Humid in the Display option the Modes graph is replaced by a graph plotting the percent relative humidity...

Page 12: ...nob to find the patient s gestational age in weeks Push the knob to enter the gestational age value and scroll down to Post natal age Push the knob to select Post natal age then turn the knob to find...

Page 13: ...rm to choose between 2 settings choose 5 to set the alarm to activate when a 0 5 C difference is read between a set temperature and the baby probe tempera ture or choose 1 0 to set the alarm to activa...

Page 14: ...0am 6 23 00 Name Birth 6 21 00 Wed Options Done Exit 85 85 set Humidifier 10 30am 6 23 00 Name Birth 6 21 00 Wed CI 31 032 CI 31 033 CI 31 029 Options Done Exit 10 30am 6 23 00 Name Birth 6 21 00 Wed...

Page 15: ...ature 10 30am 6 23 00 Name Birth 6 21 00 Wed Options Done Exit Probe2 36 8 Low Air Temperature 10 30am 6 23 00 Name Birth 6 21 00 Wed High Air Temperature Active only when the canopy is closed and the...

Page 16: ...r They are not self resetting once the alarms are activated they must be manually reset by pushing the alarm silence button even if the alarm condition is resolved Computer independent circuitry conti...

Page 17: ...en the canopy is closed the Fan Always in High Speed alarm activates when the heater temperature sensor malfunctions Air Probe Disconnect Active only when the canopy is closed the Air Probe Discon nec...

Page 18: ...Bad Membrane Switch One of the electronic switches behind the control panel buttons has failed This alarm can not be silenced Power Failure If power to the unit is cut off an audio alarm sounds and a...

Page 19: ...ty Failure Heater failed humidifier not operating Repair humidifier Options Done Exit Add Water Options Done Exit Water Reservoir Not Engaged Options Done Exit Humidity Probe Failure Options Done Exit...

Page 20: ...ST 37 C Fan Failure Fan missing 10 min 1 All Off On Off or not rotating Fan Always in Heater none 2 All On On Off High Speed temperature sensor failure Air Probe Disconnection 10 min 2 All Off On On D...

Page 21: ...n the platform None None No visual indicator Maximum Scale failure Failure None None No visual indicator Legend AST Air Set Temperature BST Baby Set Temperature Sounds 1 High priority High attention g...

Page 22: ...obes 7 Side Door Latch 8 Probe Jacks 9 Humidifier Reservoir 10 Caster Brake 11 Pleural Drainage Hanger 12 Standby Power Switch I O 13 Tubing Access Covers 1 Ventilator Slot 2 Side Door Latch 3 RS 232...

Page 23: ...E 1 Portholes 2 Control Panel 3 Upright 4 Dovetail rail 5 Canopy Raise Lower Switches 6 Pleural Drainage Hanger 7 Elevating Column 8 Bed Height Pedal 9 Raise Canopy Pedal 10 Storage Drawer 11 Side Doo...

Page 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...

Page 25: ...d from use and repaired 1 Disconnect the power cord for the mechanical portion of the preoperative checkout procedure 2 Examine the power cord for any signs of damage Replace the cord if damage is evi...

Page 26: ...r checks WARNING w w w w w Do not use the OmniBed in the presence of flammable anesthetics an explosion hazard exists under these conditions 1 Make sure the power cord is connected to the outlet on th...

Page 27: ...wer bed pedals on both sides of the unit raise and lower the bed height 11 Check the power failure alarm and the battery backed up memory Make note of the current control mode and temperature settings...

Page 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...

Page 29: ...ctrosurgical units or other instruments which radiate electrical fields can cause indirect heating by several tenths of a degree of the skin temperature probe due to absorbed electrical energy When us...

Page 30: ...raised objects could fall off and injure the operator Do not route leads or tubing over the top of the canopy or tape leads or tubing to the canopy since raising the canopy could pull leads and tubing...

Page 31: ...s Operation as a Warmer Always preheat the bed before placing the baby in the bed 1 Select a control mode manual mode is the default Preheat the bed in the manual mode Power settings in the preheat ra...

Page 32: ...id your Ohmeda Medical equipment warranty When using phototherapy lamps the probe must be directly in the path of the radiant heat of the lamp do not place the probe in an area shielded from the lamp...

Page 33: ...e side door pinch the latches together B to draw back the bolts and lower the door To slide out the mattress tray first make sure the bed is level and grasp it by one of the holes Pull the bed out tow...

Page 34: ...the bed WARNING w w w w w Before raising or lowering the bed check that there is adequate slack in tubing and leads and that no obstruc tions limit the range of motion The bed can be raised or lowered...

Page 35: ...w Remove all sources of ignition including smoking materials and sources of electrical discharge from the area when oxygen is in use In the presence of high oxygen concen trations even relatively nonf...

Page 36: ...eater cylinder not the line on the reservoir Tilt the reservoir back into place Use the Humidifier screen on the graphics display to start humidification and set a desired humidity level See accessory...

Page 37: ...formation on decoding the data stream appears in the appendix For details of the RS 232 protocol and the connector pinout refer to the Specification section WARNING w w w w w The computer or RS 232 mo...

Page 38: ...scale menu screens work Scale installation 1 Remove the clear plate and mattress from the bed 2 Fit the scale weighing platform into the recesses in the bed 3 Route the scale cable though a tubing ac...

Page 39: ...ssories WARNING w w w w w Overloading the shelves and acces sory rails can affect the stability of the unit Always try to evenly distrib ute the weight of accessories on both sides of the unit for a m...

Page 40: ...it with the nylon straps Tubing Management Arm The flexible tubing management arm aids in routing tubing to the baby It can be mounted by inserting it in any of the 4 D shaped holes located at each of...

Page 41: ...presence of oxygen will burn violently Repair Policy Warranty repair and service must be performed by an Ohmeda Medical Service Representative or at the Ohmeda Medical Service and Distribution Center...

Page 42: ...humidifier reservoir When you replace the filter mark the date on the label sup plied with the replacement filter and affix it to the filter cover panel Note This is the minimum replacement frequency...

Page 43: ...the power switch Off and unplug the unit from the power outlet 3 If the OmniBed was previously on allow it to cool for at least 30 minutes 4 Unplug the patient probe and any other probe jacks or conne...

Page 44: ...tch located on the right at the head of the bed B to release the tilt screw ball Next use the tab on right tilt pivot pin C to slide the spring loaded pin in and lift the tilt platform up out of the b...

Page 45: ...liza tion The reservoir may be steam sterilized for 3 to 5 minutes at 132 C or for 15 to 20 minutes at 121 C Many repeated sterilization cycles may lead to crazing small hairline cracks in some areas...

Page 46: ...y CAUTIONS w Use of cleaning disinfecting solutions containing chemicals not listed above i e alcohol acetone etc or chemicals in greater concentrations than those listed above may damage the probe Cl...

Page 47: ...Ammonium such as Virex Solvents such as acetone CAUTION w Use of cleaning disinfecting solutions containing chemicals not listed above or chemicals in greater concentrations than those listed above m...

Page 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...

Page 49: ...Heat reflecting probe patch 50 0203 1980 300 Mattress 6600 1116 500 Routing Clips 6 6600 0051 851 Door inner wall with portholes 6600 1201 500 Large central tubing access grommet 6600 1231 500 Small...

Page 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...

Page 51: ...C in 0 1 increments Air control temperature 20 39 C in 0 1 increments Radiant heat power 0 100 in 5 increments Humidity Servo relative humidity 30 95 in 5 increments Manual increments 1 10 Alarms High...

Page 52: ...f the matress Humidity Servo control accuracy 10 for settings up to Humidity control setting vs average 85 minimum 75 humidity at 10 cm above center of bed for settings 85 Manual humidity 75 Humidity...

Page 53: ...going warranties shall not apply if the Product has been repaired other than by Ohmeda Medical or in accordance with written instructions provided by Ohmeda Medical or altered by anyone other than Ohm...

Page 54: ...6948 Singapore Datex Ohmeda Pte Ltd 152 Beach Road 12 05 07 Gateway East Singapore 189721 Tel 65 391 8618 Fax 65 291 6618 Australia Datex Ohmeda Pty Ltd Units 1 2 149 Arthur Street P O Box 356 Homebus...

Reviews: