1-6
6600-0342-000 08/08/01
1-6
General Information
Help: Select the Help icon (question mark) to bring up the
“Help” screen. The screen will present a menu of topics on
how your OmniBed unit works. “Alarms” list all the alarm
messages on the left and more detailed descriptions of the
alarm and how to react to a given alarm on the right. “Controls”
identifies the various button and switches on the control panel.
“Operation” lists basic procedures on the left and step by step
instructions on the right. Select Done/Exit to return to the
option icon screen.
If you select the Help icon during an alarm condition, the
description of that particular alarm will automatically appear on
the screen; it is not necessary to scroll through the help menu
options to find it.
?
Alarms
Controls
Operation
Done/Exit
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
ALARMS
High Air Temp
Low Air Temp
Baby Cold
Baby Hot
Air Temp> 38C
Air Temp> 40C
down
Done/Exit
BABY COLD
1.0C under BABY SET TEMP
Temp probe poorly attached to baby
Reheating a very cold baby
BABY SET TEMP too low
Adjust and/or monitor to assure baby
temp increases.
DONE/EXIT
Baby Cold-Check Probe 1
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
OPERATION
Start Up
Incubator Use
Warmer Use
Done/Exit
1. Preheat incubator in AIR MODE
appropriately for patient. Use
COMFORT ZONE screen to calculate
an appropriate temperature.
Select temperature using the
up or down switches. Press >37C
button to adjust above 37C.
2. If BABY MODE desired, attach sensor
to baby and plug into panel.
3. Switch to BABY MODE and select
control temperature desired.
CONTINUE
DONE/EXIT
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
Alarm messages displayed here
CONTROLS
37
c
Baby Control
Override
Increase Set Point
Manual Control
Decrease Set Point
Boost Air Curtain
Air Control
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
CI.31.021
CI.31.082
CI.31.023
CI.31.024
Summary of Contents for Giraffe OmniBed
Page 1: ...GiraffeTM OmniBedTM Operator s Manual...
Page 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Page 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Page 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...