1-16
6600-0342-000 08/08/01
1-16
General Information
Alarm Table
Temperature Regulation
Alarm
Activation criteria
Alarm Silence
Audio Signal
Mode
Heater Status
Inc/Wmr
High Air Temp
1.5°C over AST
10 min
2
Air
On
On/Off
Low Air Temp
3.0°C under AST
10 min
2
Air
On
On/Off
Baby Cold
1.0°C under BST* 15 min
2
Baby
On
On/On
5 min >2°C
1
Baby Hot
1.0°C over BST*
15 min
2
Baby
On
On/On
5 min >2°C
1
Disconnected
Disconected
2 min
2
Baby
On
On/On
Baby Probe
at jack panel
1 (after 30 sec)
Air Temp >38°C
>38°C
10 min
1
All
Off
On/Off
(AST <37°C)
Air Temp >40°C
>40°C
10 min
1
All
Off
On/Off
(AST >37°C)
Fan Failure
Fan missing
10 min
1
All
Off
On/Off
or not rotating
Fan Always in
Heater
none
2
All
On
On/Off
High Speed
temperature
sensor failure
Air Probe
Disconnection
10 min
2
All
Off
On/On
Disconnect
of probe connector
1 (after 30 sec)
Air Probe Failure
Failure
10 min
2
All
Off
On/On
of compartment
air probe
1 (after 30 sec)
Power Failure
Switch on,
non-silenceable 1
All
Off
On/On
but no power
2 (after 4 sec)
System failure
any non user
non-silenceable 1
All
Off
On/On
correctable failure
Check Baby
Radiant heater at
12 min
2
Baby
Off
On/On
100% for 12 min
1 (after 15 min)
Radiant heater %
12 min
2
Manual
On
Off/On
above preheat set
1 (after 15 min)
Manual
Off
Off/On
In Transition-
Canopy stopped
10 min
2 (after 30 sec)
All
Off
On/On
Heater Off
in transition
1 (after 2 min)