4-4
6600-0342-000 08/08/01
4-4
Maintaining the OmniBed
10. Remove the mattress pad. Remove the clear plate under
the mattress. Remove the X-ray tray by sliding it all the
way out of the bed.
11. Center the translation deck and lift it up out of the chas-
sis. If the deck is pulled out toward either side, the deck
locks and can not be removed.
12. Separate the rotating bed from the translation deck by
turning it over and squeezing together the two spring
latches located in the bed’s center (A).
13. To remove the tilt platform, first slide open the finger
pocket latch, located on the right at the head of the bed
(B), to release the tilt screw ball. Next, use the tab on
right tilt pivot pin (C) to slide the spring loaded pin in and
lift the tilt platform up out of the bed.
14. Lift the pan up out of the chassis.
15. To remove the humidifier reservoir, grasp the bottom of
the reservoir and pull it away from the bed, then pull the
reservoir down away from the unit.
16. The intake filter is located behind the humidifier reservoir.
Remove the screw that holds the filter cover, then pull out
the filter to inspect or replace it.
Reverse the steps for assembly.
A
B
C
1. Mattress pad
2. Clear plate
3. X-ray tray
4. Rotating bed
5. Translation deck
6. Tilt platform
7. Pan
CI.31.059
1
2
3
4
5
6
7
A
B
C
CI.31.060
CI.31.061
CI.31.062
CI.31.063