Humidifier
Alarm
Activation criteria
Alarm Silence
Audio Signal
Add Water
Reservoir low on water
10 min
3
Check Water Reservoir
Reservoir missing or not
10 min
3
fully inserted
Humidity Probe Failure
Probe failed
10 min
1
open or shorted
Humidifier Failure
Failure
None
1
X appears on icon
Scale
Weight Exceeds
> 8 kg on the platform
None
None
No visual indicator
Maximum
Scale failure
Failure
None
None
No visual indicator
Legend
AST = Air Set Temperature
BST = Baby Set Temperature
Sounds
1 = High priority - High attention getting - e.g. two tone siren
2 = Low priority - Alert, but not requiring immediate attention - e.g. soft repeating tones with 15 to 60 seconds between
soundings.
3 = Informational - Not necessarily requiring attention or not immediately - e.g. soft repeating tones with 1-5 minutes between
soundings.
*
Activation criteria can be changed to 0.5°C by the user from the Set Up/Customize screen.