Operating the OmniBed
w
w
w
w
w
When operating as a radiant warmer, use the
baby mode unless the manual mode is
specifically prescribed. While both modes
require patient monitoring, the manual mode
requires constant attention. In the manual
mode, you must take the responsibility for
detecting changes in the environment
(drafts, direct sunlight, phototherapy lamp
usage, etc.) or the patient condition requiring
heater adjustments in response to these
changes. In the baby mode, the warmer
automatically adjusts heater output to
maintain the desired skin temperature,
reducing (but not eliminating) the need to
monitor the patient and make adjustments to
the equipment.
w
w
w
w
w
Using an OmniBed with a fan motor that has
worn parts can produce unacceptable noise
levels within the infant compartment.
w
w
w
w
w
Direct sunlight or other radiant heat sources
can cause an increase in OmniBed tempera-
ture to dangerous levels.
w
w
w
w
w
Do not leave the baby unattended while any
side panels are lowered or removed.
w
w
w
w
w
Do not place objects on top of the canopy.
When the canopy is raised, objects could fall
off and injure the operator. Do not route
leads or tubing over the top of the canopy or
tape leads or tubing to the canopy, since
raising the canopy could pull leads and
tubing from patient or equipment connec-
tions. Always check for accessories or
objects that could be in the canopy’s path
before raising or lowering the canopy.
CAUTIONS
w
w
w
w
w
Proper temperature control depends on continu-
ous, unobstructed air circulation. Do not cover
air circulation openings around the bed as
obstruction will result in loss of air circulation,
loss of heat, and carbon dioxide buildup.
w
Do not mount or rest a radiant warmer or
incandescent light on or over the OmniBed;
ineffective heating and damage to the canopy
may result. If using incandescent phototherapy
lamps, check manufacturer’s recommended
minimum distance to equipment.
1. Plug the unit into a properly rated AC power outlet and set the
caster brakes.
2. Place the mains power switch, located by the outlets on the
back of the unit, in the On (I) position.
3. Place the power Stand By switch, located on the left side of the
jack panel, in the On (I) position.
4. Depending on whether the unit is to be used as an incubator or
as a warmer, raise or lower the canopy using the hand switches
located on either side of the bed near the uprights.