1-8
6600-0342-000 08/08/01
1-8
General Information
Comfort zone: Select Comfort Zone icon (thermometer and
face) to bring up the “Comfort Zone” screen.
Important: The comfort zone screen is only a reference
information screen used for calculating suggested temperature
ranges* during air control mode from data you supply. You
must manually enter the temperature settings using the
buttons on the temperature regulation controls.
Push the knob to select weight then enter the patient’s weight
range in grams by turning the knob in 100 gram increments.
Push the knob to enter the weight value.
Scroll to Gest age.
Push the knob to select Gest age then turn the knob to find the
patient’s gestational age in weeks.
Push the knob to enter the gestational age value and scroll
down to Post natal age.
Push the knob to select Post natal age then turn the knob to
find the age in days. Push the knob to enter the postnatal age.
The suggested incubator air set temperature range will now
appear in degrees Celsius. If you wish to use this suggested
temperature setting, enter it using temperature control buttons
to the left of the screen.
Select done/exit to return to the option menu screen.
* Adapted from Sauer, Dane and Visser, Archives of Disease in Childhood, 1984. 59, 18-22.
?
Options
Done/Exit
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
?
Options
Done/Exit
Comfort Zone
Weight
Gest age
Post natal age
Done/Exit
Suggested Temp Range
34.4 - 35.4 C
1400-1500 grams
34 wks
1 day
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
CI.31.027
CI.31.028
Summary of Contents for Giraffe OmniBed
Page 1: ...GiraffeTM OmniBedTM Operator s Manual...
Page 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Page 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Page 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...