Operating your Moisture Analyzer - Section 6
Using the arrow buttons, scroll to MG, now enter the weight loss (1 mg to
10 mg).
AUTO FREE
AUTO FREE
SWITCHOFF AT LESS THAN
5 MG IN
60
SECONDS
Using the arrow buttons, enter the time (10 seconds to 120 seconds).
Press the
Enter
button.
TEST PARAMETERS
Using the arrow buttons, scroll to DISPLAY and press the
Enter
button.
6.6 Result
Displayed result offers you the choice of whether your display indicates % moisture, % solids, % regain, grams
or custom unit during the measurement process
TEST PARAMETERS
Using the arrow buttons, scroll to either % MOISTURE, % SOLIDS, %
REGAIN, GRAMS or CUSTOM UNITS and press the
Enter
button.
NOTES
: 1. If you change the displayed result during a test, the final result
and the printed results will be in displayed units.
2.
When CUSTOM is selected, the parameters for custom units
are changed in paragraph 6.7.
EXIT WITHOUT SAVING
EXIT WITHOUT SAVING
EXIT WITHOUT SAVING
SWITCH OFF AT LESS THAN
5
MG IN 60 SECONDS
TEST ID:
DEFAULT
DRY TEMP:
125C
PROFILE:
STANDARD
SWITCH OFF: AFREE
RESULT:
MOISTURE%
CUSTOM :
OFF
% MOISTURE
6.7 Custom units
Custom unit allows the creation of your own custom weighing unit. It permits entering a conversion factor
which the Moisture Analyzer will use to convert grams to the desired unit of measure. The custom unit is
available under the Test Parameters menu, Display settings. When the Moisture Analyzer is completely set up
and running a test, you have the option under the Display to view % Moisture, % Solids, % Regain, Grams or
Custom Unit.
Conversion Factor x Weight in grams = Weight in custom unit
Conversion factors are expressed in scientific notation and entered into the balance in three parts:
1) a number between 0.1 and 1.999999 called the mantissa
2) a power of 10 called the exponent
3) a least significant digit (LSD)