Getting to know your Moisture Analyzer - Section 1
Your Moisture Analyzer employs state of the art technology and meets the latest de-
mands regarding instrument safety. Improper operation can endanger personnel and
can cause property damage. For safe and dependable operation, please comply with
the following instructions:
The Moisture Analyzer is used for determination of the moisture in samples. Please
use the instrument exclusively for this purpose. Any other type of use can endanger
personnel and damage the instrument or other property.
The Moisture Analyzer must not be operated in a hazardous environment and only
under ambient conditions specified in these instructions.
The Moisture Analyzer may be operated only by trained personnel who are familiar
with the properties of the samples used and with the handling of the instrument.
Your Moisture Analyzer is supplied with a 3-pin power cable with an equipment
grounding conductor. Only extension cables which meet the relevant standards and
also have an equipment grounding conductor may be used. Intentional disconnec-
tion of the equipment grounding conductor is prohibited.
1.4 Safety is first
The Halogen Moisture Analyzer works with heat!
Ensure sufficient free space around the instrument to avoid heat accumulation and
overheating (approximately 1 m free space above the instrument).
Never place flammable materials on, below or next to the instrument as the area
around the dryer unit warms up.
Exercise caution when removing the sample. The sample itself, the sample cham-
ber and any sample containers may still be very hot.
During operation, you should never open the dryer unit as the ring-shaped heating
element or its protective glass can reach 400°C! If you have to open the dryer unit on
occasion, disconnect the instrument from the power supply and wait until the dryer
unit has cooled down completely.