Section 6 - Operating your Moisture Analyzer
6.3 Setting the drying profile
The drying profile contains four settings, Standard, Fast, Ramp and Step. Section 6, paragraph 6.4 describes
each of the drying profiles. Review the profile descriptions. Depending upon the sample material to be analyzed,
select the appropriate profile and proceed as follows:
Standard profile
Using the arrow buttons, scroll to PROFILE and press the
Enter
button.
DRYING PROFILE
Using the arrow buttons, scroll to STANDARD and press the
Enter
button.
Temperature profile
Final
temperature
Ambient
temperature
Standard
Time
Start
EXIT WITHOUT SAVING
TEST PARAMETERS
TEST ID:
-DEFAULT-
PROFILE:
STANDARD
DRY TEMP:
180C
SWITCH OFF: AUTO 60
RESULT:
MOISTURE %
CUSTOM:
OFF
STANDARD
Fast profile
Using the arrow buttons, scroll to PROFILE and press the
Enter
button.
DRYING PROFILE
EXIT WITHOUT SAVING
Using the arrow buttons, scroll to FAST and press the
Enter
button.
Temperature profile
Final
temperature
Ambient
temperature
Fast drying
Time
Start
TEST PARAMETERS
TEST ID:
-DEFAULT-
PROFILE:
STANDARD
DRY TEMP:
180C
SWITCH OFF: AUTO 60
RESULT:
MOISTURE %
CUSTOM:
OFF
FAST