2 6
Analyzer setup - Section 5
5.3 Temperature calibration
You must have a temperature calibration kit to perform this procedure. If the Moisture Analyzer has been re-
cently used, allow at least 30 minutes before performing calibration.
Press the
Setup
button. Using the arrow buttons, scroll to TEMP CAL.
Press the
Enter
button. You are now prompted to remove the pan
handler and pan support. Replace the pan handler and place a tem-
perature calibration unit on the pan handler.
Note
: The unit
will not calibrate
with the pan support in place.
ANALYZER SETUP
LOCK OUT
WEIGHT CAL
TEMP CAL
LANGUAGE
BEEPER
TIME-DATE
Note
: The temperature calibration kit is available as optional equipment.
TEMP CAL
Press the
Enter
button to initiate the temperature calibration process.
Follow the screen prompts throughout the process.
TEMP CAL 100C
The dryer unit is heated to a temperature of 100°C. You can observe the
progress on the display as the dryer temperature and count down period
are displayed. After 15 minutes, read the thermometer through the
inspection window on the cover and enter this temperature. Using the
up/down arrow buttons, adjust the display reading to agree with the
thermometer. Use the left/right arrow buttons to highlight ACCEPT NEW
CAL, then press the
Enter
button. You have 10 minutes to make this
adjustment, otherwise the calibration is terminated.
TEMP CAL 160C
This is a two point adjustment (100°C and 160°C). The dryer unit now
heats to the second temperature (160°C). Adjustment of the temperature
is defined by two points. Proceed exactly as you did for the first tempera-
ture. After you have set the display to agree with the thermometer,
highlight ACCEPT NEW CAL, then press the
Enter
button. The display
returns to ANALYZER SETUP.
WARNING
: Be careful when removing the temperature calibration unit
from the dryer unit as it can be very hot. Allow it to cool down by open-
ing the cover before removing.
Remove the calibration unit. Replace the pan support and the pan
handler in their proper positions.
CURRENT TEMP:
--- C
TM TO CAL POINT:
15:00 MIN
ADJ CAL READING:
100
C
CURRENT TEMP:
100 C
TM TO CAL POINT:
15:00 MIN
ADJ TO CAL READING:
160
C
REMOVE PAN HANDLER
AND PAN SUPPORT
ABORT CALIBRATION
ACCEPTS NEW CAL
ABORT CALIBRATION
ACCEPTS NEW CAL
ABORT CALIBRATION