Installation - Section 2
2.2 Selecting the location
The Moisture Analyzer should always be used in an environment which is free from excessive air currents, corro-
sives, vibration, and temperature or humidity extremes. These factors will affect displayed weight readings.
DO NOT
install the Moisture Analyzer:
Next to open windows or doors causing drafts or
rapid temperature changes.
Near air conditioning or heat vents.
Near vibrating, rotating or reciprocating equipment.
Near magnetic fields or equipment that generate
magnetic fields.
On an unlevel work surface.
In confined areas, allow sufficient space around the
instrument for ease of operation and keep away from
radiating heat sources.
On/Off
Test Menu
Display
Setup
Enter
TARE
STAR
T/ST
OP
MB45
On/Off
Test Menu
Display
Setup
Enter
TARE
STAR
T/ST
OP
MB45
On/Off
Test M
enu
Display
Setup
Enter
TARE
STAR
T/ST
OP
MB45
Lift the cover straight up and Install the heat shield in the base of the
heating chamber.
Install the draft shield element (only one position possible) on top of the
heat shield.
Install the pan support into postion. Turn the pan support until it en-
gages. In the locked position, the rear arm of the pan support points
directly towards the rear of the analyzer.
2.3 Installing the heat shield, draft shield and sample pan holder
PAN SUPPORT
DRAFT SHIELD
HEAT SHIELD