Section 2 - Installation
2.5 Connecting to a power supply
Check to ensure the voltage identification label on the Moisture Analyzer
matches your local line voltage. If this is not the case, do not connect the
Moisture Analyzer to the power supply, contact your responsible OHAUS
dealer.
The halogen dryer unit is designed to operate at a specific line voltage
(120 V ac or 240 V ac). The dryer unit is installed at the factory and is
matched to the particular line voltage of the country of destination.
Connection to a line voltage that is too high can lead to burning out the
halogen heater, whereas, a supply voltage that is too low will prolong the
drying process and the instrument may not operate properly.
2.4 Leveling the Moisture Analyzer
Exact horizontal positioning and stable installation are prerequisites for repeatable results. To compensate for
small irregularities or inclinations at the location, the instrument can be leveled.
Adjust the leveling feet at the rear of the Moisture Anlyzer until the air
bubble in the indicator is centered. The level indicator is located under
the cover towards the rear of the Moisture Analyzer.
NOTE
: The instrument should be leveled each time its location is
changed.
Leveling
Foot
Leveling
Indicator