EN-84
EXPLORER BALANCES
4.13
Pipette Adjustment
(not available in EX12001, EX24001, EX35001)
Pipette adjustment checks the accuracy and precision values of pipettes by weight analysis. The balance has a
built-in density reference table for water at temperatures between 10°C and 30°C. If other liquids are used for
pipette calibration, enter the liquid’s density at current room temperature in g/cm
3
. Since all calculations are made
within the balance, the atmospheric pressure is also required.
Pipette adjustment can be accomplished
manually
(with a key press after each step) or
automatically
(weights
are automatically recorded when stable). Calculations made within the balance provide results of inaccuracy and
imprecision. Up to 30 samples can be used.
The
PIPETTE ADJUSTMENT
Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions Application Icon
4.13.1 Pipette Adjustment
– Manual
(default)
Confirm the default values displayed are correct:
Nominal, Inaccuracy, Imprecision, Density, Atmospheric
Pressure, Water Temperature, etc.
To edit the default values, touch the
Edit Settings
button.
The
Edit Settings
Screen.
Settings available:
Nominal Capacity, Water Temp,
Inaccuracy, Atmospheric Press,
Imprecision, Pipette Name,
Pipette Number
Functions Available:
Return to Application
To adjust the pipette Nominal Capacity value,
touch the
Nominal
button.
Enter the Nominal Capacity in the keyboard screen that
appears, and press
.
The display returns to the previous screen with the new value
highlighted.
Note:
Capacity can be measured in
mL
or
uL
, depending on the
Application Setup.
www.
.com
1.800.561.8187