TABLE 5-1 Calibration Masses
Model
Span Calibration Points
Weight Class
EX125D; EX125DM
50g, 100g
ASTM Class 1
OIML E2
EX125; EX125M
50g, 100g
ASTM Class 1
OIML E2
EX225D; EX225DM
100g, 150g, 200g
ASTM Class 1
OIML E2
EX225D/AD;
EX225DM/AD
100g, 150g, 200g
ASTM Class 1
OIML E2
EX225/AD; EX225M/AD
100g, 150g, 200g
ASTM Class 1
OIML E2
EX124
50g,
100g
ASTM Class 1
OIML E2
EX224
100g, 150g,
200g
ASTM Class 1
OIML E2
EX324, M, N
150g, 200g,
300g
ASTM Class 1
OIML E2
EX223
100g, 150g,
200g
ASTM Class 1
OIML E2
EX423
200g, 300g,
400g
ASTM Class 1
OIML E2
EX623
300g, 400g,
500g
,
600g
ASTM Class 1
OIML E2
EX1103, M, N
500g,
1000g
ASTM Class 1
OIML E2
EX2202
1000g,
2000g
ASTM Class 1
OIML E2
EX4202
2000g, 3000g,
4000g
ASTM Class 1
OIML E2
EX6202
3000g, 4000g,
5000g
, 6000g
ASTM Class 1
OIML E2
EX10202, M, N
5000g,
10,000g
ASTM Class 1
OIML E2
EX6201
3000g, 4000g,
5000g
, 6000g
ASTM Class 2
OIML F1
EX10201
5000g,
10,000g
ASTM Class 1
OIML E2
EX12001
3000g,5000g,10000g,
12000g
ASTM Class 2
OIML F1
EX24001
10000g,15000g,20000g,
24000g
ASTM Class 1
OIML F1
EX35001
10000g,20000g,30000g,
35000g
ASTM Class 1
OIML F1
With the balance turned ON and no load on the pan, touch Span Calibration to initiate the procedure.
Additional calibration values to be used are shown on the display. The best accuracy is achieved using
the mass closest to the full span value.
Note:
To change the span calibration point, touch the alternate weight shown on the display.
Follow the screen instructions and place the specified calibration weight on the scale when prompted to do so.
When complete, the display shows the Span calibration status and returns to the current application.
5.3.7
User Calibration
User Calibration is used to calibrate the balance using a user defined mass.
With no load on the scale, press
User Calibration
from the calibration sub-menu.
The display shows the current full load (span) calibration point and calibration unit of measure.
To change the value to a user-defined weight, touch
ENTER VALUE
.
Enter the weight via the popup screen. The balance then prompts for the User-defined weight.
Place the specified calibration weight on the balance and the calibration process completes.
The display shows status, then returns to the current application.
5.3.8
Calibration Test
Use Calibration Test to compare a known calibration weight against the stored span calibration data.
With no load on the scale, press
Cal Test
to begin the process.
The balance captures the zero point, then prompts for the span value.
The display shows status, followed by the difference between the current calibration weight and the
stored calibration data.
After 3 seconds, the display returns to the current application.