EXPLORER BALANCES
EN-99
5.3 Calibration
Explorer Balances offer a choice of six calibration methods: Internal Calibration, Automatic Calibration
(AutoCal), Calibration Adjust, Span Calibration, User Calibration and Calibration Test.
ExCal models have three calibration methods: Span Calibration, User Calibration and Calibration test.
Note:
Do not disturb the balance during any calibration.
5.3.1
Calibration sub-menu
AutoCal™
Internal
Calibration
Automatic
Calibration
Auto
TM
Interval
(Hours)
AutoCal™
Adjustment
Span
Calibration
User
Calibration
Calibration
Test
5.3.2
Internal Calibration (AutoCal
™)
Calibration is accomplished with the internal calibration mass. Internal calibration can be performed at
any time, provided the balance has warmed up to operating temperature and is level.
With the Balance turned
ON
and no load on the pan,
touch
AutoCal™ Internal Calibration
, and then touch
Calibration
.
The Balance begins to calibrate.
Alternately, touch the shortcut
Calibration
Icon from any Application home screen
to initiate an Internal Calibration.
Note:
For Non-Approved models, when AutoCal Internal Calibration is turned off, it changes to span calibration.
For Approved models, when AutoCal Internal Calibration is turned off, it is hidden.
The display shows the status, then returns to the current application.
To cancel at any time, press
Cancel
.
5.3.3
Automatic Calibration
When
Automatic Calibration
is set ON, the balance performs a self calibration:
• when it senses a temperature change of 1.5°C
•
or every 11 hours
AutoCal will automatically calibrate the Balance (using the internal mass) each time there is a change in
temperature significant enough to affect accuracy.