EN-34
EXPLORER BALANCES
4.2.3
Parts Counting
– Check
This feature permits establishing a set quantity of pieces as criteria for similar items that can be quickly checked
against a sample.
Press the
Setup
(wrench)
button to enter
the Counting
Setup menu.
Change
Counting
Mode to
Check
Counting
.
The
PARTS COUNTING
– Check
Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions
The default (or last) Average Piece Weight (APW) is
displayed. Place objects on the pan to display the
Check Status
(number of pieces).
4.2.3.1 Edit Settings:
To use the Parts Counting-Check mode,
the
Over Limit
and
Under Limit
values must be set.
The
Average Piece Weight
and standard sample size can also
be set. Follow these steps:
Touch the
Edit Settings
button. The Settings screen appears.
The buttons for
APW
,
Samples, Sample Name, Sample ID,
Batch ID, Clear All.
Note:
please refer to section 4.1.1 for details about setting
Sample Name, Sample ID, Batch ID and Clear All.
Over Limit
and
Under Limit
show default (or previously set)
values. You can also touch
Batch ID
to set an ID for your
samples. To change any value, touch the button. A keyboard
screen appears.
(See keyboard screens, next page.)
To set a new average piece weight, press the
Recalculate APW
button.
A new screen appears with the message:
Place reference weight on the pan. Press Accept to
continue.
After following these instructions, press
Return to Application
on the main
Edit Settings
screen.
www.
.com
1.800.561.8187