Clamp pincer CP 10 / CP 20 / CC 20
7
Description of PC software
The initial calibration closing force for the various devices is factory-set as follows:
•
CP 10 =
Closing force 1000 N
•
CP 20 =
Closing force 3500 N
•
CC 20 =
Closing force 3500 N
Closing force calibration
Closing force calibration is performed with the target value for the closing force, so as to achieve the highest
accuracy by the pincer system.
Five measured closures are performed during the course of the closing force calibration. An average is
calculated from the results of these five measured closures, and this is loaded to the device.
7.5 Firmware
menu
New firmware is loaded on the device with the Firmware menu (see chapter 4.4 Updating firmware, p. 4-15).
Fig. 34:
Firmware menu
Issue 11.15
08902961
7-43