Clamp pincer CP 10 / CP 20 / CC 20
5
Use of clamp pincer
Closing force calibration
INFO
Reset the measuring instrument before starting the closing force calibration.
If the CAL01 is being used, select the "Average” setting.
After the measured value for the initial calibration has
been sent, the “Calibration / closing force-calibration“
window will open.
The closing force target value will already be entered
and doesn’t need to be re-entered.
NOTICE
The target closing force is carried over from the
"Closing force [N]" input field of the "Closing data"
tab (see chapter 7.3, p. 7-40):
1. Click on the “Start closing force calibration”
button.
The button lights up green and the “Closing force
calibration” information window opens. This
displays the next steps and contains a counter
showing the number of measurement closures.
The counter provides an indication of the number
of closures performed during calibration.
Once the 5 closures required for calculating the
average have been performed, the “Closing
force-calibration” indicator closes automatically.
2. Enter the average of the measuring values (e.g.
from CAL01) into the “Average value [N]” field in
the “Calibration values” area.
The “Send average value to device” button lights
up green.
3. Press the “Send average value to device” button.
The measured value is acknowledged and sent
to the device.
Issue 11.15
08902961
5-22