OM0301
August 20
Page 88 of 105
•
Fan Relay interlock:
This parameter allows for the selection of the fan relay interlock operation. When
the fans relay interlock is on, and the blower switch is pending or blue, the fan relay will be off or not
energized. If the fan relay interlock is off and the blower switch is pending or blue, the fan relay will be
on or energized. In either case the fan relay will be on when the blower switch is on or green and off
when the blower switch is off or not colored.
•
Blower FL Light Interlock:
This parameter allows for the selection of the FL light option to be
interlocked to the blower. When the blower FL light interlock is on, the FL light operation will be
interlocked to the blower. When the blower FL light interlock is off, the FL light can be turned on at any
time.