OM0301
August 20
Page 76 of 105
8.2.2.1 Blower/Damper Setup
Blower Options
The blower options menu allows
A QUALIFIED TECHNICIAN
to run the cabinet in manual mode. This means with
no controls or alarms activated. Manual Blower control: When the manual blower control is on, the downflow and
inflow displays will indicate
Manual Blower Control
This parameter allows
ONLY THE CABINET TECHNICIAN
to run the cabinet in manual mode. This means with no
controls or alarms activated. When the manual control is on, the downflow and inflow displays will indicate
nominal setpoints. Airflow adjustments can be made in the manual mode by going into airflow calibration and
adjusting the blower duty cycle. The blower duty cycle will remain constant in manual mode. The display will also
indicate the manual control is activated. When the manual control is off, full automatic control resumes.
• Blower Lockout
This parameter allows the access to turn the blower on or off to be restricted by the use of a password.
When the blower lockout is on, pressing the blower icon will produce a numerical password screen.
When the blower lockout is off, the blower may be turned on and off without restriction.
Summary of Contents for LabGard ES NU-565-400
Page 7: ...OM0301 August 20 Page 7 of 105...
Page 9: ...OM0301 August 20 Page 9 of 105...
Page 10: ...OM0301 August 20 Page 10 of 105...
Page 11: ...OM0301 August 20 Page 11 of 105...
Page 17: ...OM0301 August 20 Page 17 of 105...
Page 18: ...OM0301 August 20 Page 18 of 105...
Page 45: ...OM0301 August 20 Page 45 of 105...
Page 50: ...OM0301 August 20 Page 50 of 105...
Page 82: ...OM0301 August 20 Page 82 of 105...
Page 98: ...OM0301 August 20 Page 98 of 105...
Page 99: ...OM0301 August 20 Page 99 of 105...
Page 100: ...OM0301 August 20 Page 100 of 105...
Page 101: ...OM0301 August 20 Page 101 of 105...
Page 102: ...OM0301 August 20 Page 102 of 105...
Page 103: ...OM0301 August 20 Page 103 of 105...
Page 104: ...OM0301 August 20 Page 104 of 105...
Page 105: ...OM0301 August 20 Page 105 of 105...