7.10.4 Cabinet Reset
The main control board has two software operating resets available for qualified service personnel.
The two types are the following:
Factory Reset - Resets setpoints and selected option settings. Factory reset should be used in the event the
system memory develops an error in operation. Cabinet type, motor type and calibration data
will not be affected with this reset.
Master Reset - Resets all calibration, cabinet type, motor type, sensor data, and options settings back to
default settings. Master Reset should only need to be used for a main control board
replacement.
After pressing the Calibration/Service menu item, a password screen will appear. The default password is “9876”.
Once the service password is entered, the Calibration/Service menu will appear.
Select CABINET RESET from the menu.
Summary of Contents for LabGard ES NU-565-400
Page 7: ...OM0301 August 20 Page 7 of 105...
Page 9: ...OM0301 August 20 Page 9 of 105...
Page 10: ...OM0301 August 20 Page 10 of 105...
Page 11: ...OM0301 August 20 Page 11 of 105...
Page 17: ...OM0301 August 20 Page 17 of 105...
Page 18: ...OM0301 August 20 Page 18 of 105...
Page 45: ...OM0301 August 20 Page 45 of 105...
Page 50: ...OM0301 August 20 Page 50 of 105...
Page 82: ...OM0301 August 20 Page 82 of 105...
Page 98: ...OM0301 August 20 Page 98 of 105...
Page 99: ...OM0301 August 20 Page 99 of 105...
Page 100: ...OM0301 August 20 Page 100 of 105...
Page 101: ...OM0301 August 20 Page 101 of 105...
Page 102: ...OM0301 August 20 Page 102 of 105...
Page 103: ...OM0301 August 20 Page 103 of 105...
Page 104: ...OM0301 August 20 Page 104 of 105...
Page 105: ...OM0301 August 20 Page 105 of 105...