OM0301
August 20
Page 70 of 105
•
Once factory reset is complete, return to Calibration/Service menu to enter any options.
Once the MASTER RESET icon is pressed, the display screen will remain the same for approximately a few seconds.
Once the reset process is complete the display screen will revert back to the Nuaire logo main menu. At this point
the cabinet MUST be turned off to complete the process. Either unplug the cabinet or use the power switch within
the control center to turn off the cabinet. Turn the cabinet back on the display screen will remain blank for up to a
minute, then will indicate “Power Loss Alert”, press the screen to clear the message and return to the
Calibration/Service menu to enter cabinet type and perform airflow calibration.