OM0301
August 20
Page 77 of 105
•
Password Reset
The default password is “1234” and may be changed using the password reset.
Night Setback Calibration (see calibration section)
The night set back calibration is performed within the calibration mode after the inflow and downflow
calibration procedure is complete.
Exhaust Damper Set up
Exhaust Airflow Control
This parameter allows for the selection of time, programmable from 30 to 240 seconds
determining how often the exhaust controller updates the control signal to the automatic damper (i.e. NU-951-
012) if installed. Reducing the time will cause the damper to react quicker to change but may result in control
Overshoot and oscillations. Increasing the time will cause the damper to react slower to changes but may not keep
up with normal systems fluctuations. Depending upon the HVAC system, changing this parameter allows control
flexibility for exhaust system optimization.
Summary of Contents for LabGard ES NU-565-400
Page 7: ...OM0301 August 20 Page 7 of 105...
Page 9: ...OM0301 August 20 Page 9 of 105...
Page 10: ...OM0301 August 20 Page 10 of 105...
Page 11: ...OM0301 August 20 Page 11 of 105...
Page 17: ...OM0301 August 20 Page 17 of 105...
Page 18: ...OM0301 August 20 Page 18 of 105...
Page 45: ...OM0301 August 20 Page 45 of 105...
Page 50: ...OM0301 August 20 Page 50 of 105...
Page 82: ...OM0301 August 20 Page 82 of 105...
Page 98: ...OM0301 August 20 Page 98 of 105...
Page 99: ...OM0301 August 20 Page 99 of 105...
Page 100: ...OM0301 August 20 Page 100 of 105...
Page 101: ...OM0301 August 20 Page 101 of 105...
Page 102: ...OM0301 August 20 Page 102 of 105...
Page 103: ...OM0301 August 20 Page 103 of 105...
Page 104: ...OM0301 August 20 Page 104 of 105...
Page 105: ...OM0301 August 20 Page 105 of 105...