5.4 Certification Testing Methods and Equipment
After installation and prior to use, Nuaire recommends that the cabinet be certified or commissioned to factory standards.
As a part of certification, the certifier should go through the following initial checklist to assure all aspects of the LFBSC
installation are complete and ready for certification.
•
Review product installation
−
Exhaust connection
−
Damper valve installed correctly with label toward front
−
LFBSC base stand level
•
Verify airflow sensor shroud is in place
−
Downflow
•
Verify configuration type selection for specific model * (see section 7.5.2)
•
Verify setpoints and alarm limits for specific model * (see section 7.5.2)
•
Perform BCS certification
−
At a minimum, the following tests should be performed:
HEPA filter leak test
Downflow velocity test
Inflow velocity test
Airflow smoke patterns
Site installation assessment tests
•
Perform Site Assessment Tests
−
The NU-565 requires verification of the supply fan interlock and back-up pressure switch operation utilizing
independent exhaust volume measurement instrument (DIM). Per NSF/ANSI 49, a 20% loss of exhaust volume
must produce an airflow alarm within 15 seconds.
The testing methods and equipment required are specified on the factory inspection report included with this manual.
(See insert in back cover)
NOTE: IT IS RECOMMENDED THAT THESE TESTS BE PERFORMED BY A QUALIFIED TECHNICIAN WHO IS
FAMILIAR WITH THE METHODS AND PROCEDURES FOR CERTIFYING BIOSAFETY FUME HOOD/CABINETS
(SEE INSERT)
NOTE: AFTER THE INITIAL CERTIFICATION, NUAIRE RECOMMENDS THAT THE CABINET BE RECERTIFIED (AT A
MINIMUM) ON AN ANNUAL BASIS AND AFTER EVERY FILTER CHANGE OR MAINTENANCE ACTION OR
ANY TIME THE OPERATOR FEELS IT IS NECESSARY.
Note that the LABGARD® ES cabinets, filters, and seals provide premium performance. Quality Control in both design and
manufacturing assure superior reliability. However, protection to both product and operator is so vital that certification to
the performance requirements should be accomplished as stated to ensure Biosafety established by the factory standards.
* If the specific model is a special product with non-standard setpoints and alarm limits, the new values will
be located on the factory Inspection Report.
Summary of Contents for LabGard ES NU-565-400
Page 7: ...OM0301 August 20 Page 7 of 105...
Page 9: ...OM0301 August 20 Page 9 of 105...
Page 10: ...OM0301 August 20 Page 10 of 105...
Page 11: ...OM0301 August 20 Page 11 of 105...
Page 17: ...OM0301 August 20 Page 17 of 105...
Page 18: ...OM0301 August 20 Page 18 of 105...
Page 45: ...OM0301 August 20 Page 45 of 105...
Page 50: ...OM0301 August 20 Page 50 of 105...
Page 82: ...OM0301 August 20 Page 82 of 105...
Page 98: ...OM0301 August 20 Page 98 of 105...
Page 99: ...OM0301 August 20 Page 99 of 105...
Page 100: ...OM0301 August 20 Page 100 of 105...
Page 101: ...OM0301 August 20 Page 101 of 105...
Page 102: ...OM0301 August 20 Page 102 of 105...
Page 103: ...OM0301 August 20 Page 103 of 105...
Page 104: ...OM0301 August 20 Page 104 of 105...
Page 105: ...OM0301 August 20 Page 105 of 105...