UV Window interlock
:
This parameter allows for the selection of the window closed switch to be
interlocked with the UV light option. When the window interlock is on, the window must be closed for
the UV light to operate. When the window interlock is off the UV light can be turned on regardless of
window position.
NOTE
: In addition to the
TOUCHLINK
system UV window interlock there is a double redundant UV window
interlock relay. To override the UV window interlock for service purposes only, both interlocks must be changed
through the
TOUCHLINK
system and shorting the relay connection (see electrical schematic for reference).
•
Blower Window Interlock
: This parameter allows for the selection of the window closed switch to be
interlocked to the blower. When the blower interlock is on, the blower will turn off when the window is
closed. When the blower is off, the blower will continue to run when the window is closed
Summary of Contents for LabGard ES NU-565-400
Page 7: ...OM0301 August 20 Page 7 of 105...
Page 9: ...OM0301 August 20 Page 9 of 105...
Page 10: ...OM0301 August 20 Page 10 of 105...
Page 11: ...OM0301 August 20 Page 11 of 105...
Page 17: ...OM0301 August 20 Page 17 of 105...
Page 18: ...OM0301 August 20 Page 18 of 105...
Page 45: ...OM0301 August 20 Page 45 of 105...
Page 50: ...OM0301 August 20 Page 50 of 105...
Page 82: ...OM0301 August 20 Page 82 of 105...
Page 98: ...OM0301 August 20 Page 98 of 105...
Page 99: ...OM0301 August 20 Page 99 of 105...
Page 100: ...OM0301 August 20 Page 100 of 105...
Page 101: ...OM0301 August 20 Page 101 of 105...
Page 102: ...OM0301 August 20 Page 102 of 105...
Page 103: ...OM0301 August 20 Page 103 of 105...
Page 104: ...OM0301 August 20 Page 104 of 105...
Page 105: ...OM0301 August 20 Page 105 of 105...