OM0301
August 20
Page 60 of 105
back ON/OFF. Press home icon the save screen pop up, press save to enter selection.
Step 2: Select CALIBRATE NIGHT SETBACK
Step 3: At this point, the night setback is active and the butterfly valve will begin to close. Now measure
the inflow volume using a Direct Reading Instrument of inflow velocity measurement method.
The goal is to reduce the inflow volume to the following values, which represents approximately
100 fpm (.51m/s) inflow, or may be set at any other values within the damper operational range
of 30% to 80%.
NU-565-400
270 CFM/459 CMH
NU-565-600
410 CFM/697 CMH
Please note, in the night setback operation, the electronic airflow control system will cause the
butterfly valve actuator to hunt more operating at lower airflow volumes. When taking the
inflow volume measurements, average several readings to obtain the most accurate results.
Summary of Contents for LabGard ES NU-565-400
Page 7: ...OM0301 August 20 Page 7 of 105...
Page 9: ...OM0301 August 20 Page 9 of 105...
Page 10: ...OM0301 August 20 Page 10 of 105...
Page 11: ...OM0301 August 20 Page 11 of 105...
Page 17: ...OM0301 August 20 Page 17 of 105...
Page 18: ...OM0301 August 20 Page 18 of 105...
Page 45: ...OM0301 August 20 Page 45 of 105...
Page 50: ...OM0301 August 20 Page 50 of 105...
Page 82: ...OM0301 August 20 Page 82 of 105...
Page 98: ...OM0301 August 20 Page 98 of 105...
Page 99: ...OM0301 August 20 Page 99 of 105...
Page 100: ...OM0301 August 20 Page 100 of 105...
Page 101: ...OM0301 August 20 Page 101 of 105...
Page 102: ...OM0301 August 20 Page 102 of 105...
Page 103: ...OM0301 August 20 Page 103 of 105...
Page 104: ...OM0301 August 20 Page 104 of 105...
Page 105: ...OM0301 August 20 Page 105 of 105...