MicroE Optira Installation Manual
Celera Motion
IM-1001 | REV181211
Page 34
©2018 Celera Motion
11.3.2 Install Tape Scale Using the Applicator Tool
Use the steps in this section to install Optira Tape Scales using the Applicator Tool.
The Applicator Tool
Model PILT-AT
The Applicator Tool is recommended for tape scale lengths greater than 300 mm, but can be
used for shorter lengths. It is installed in the sensor mounting holes to slide along the motion axis
while feeding the tape scale to attach to a mounting surface.
Tape Travel Path
The path for the tape scale through the applicator tool is shown in the following diagram.
Left end of tool
Roller/Dowel
Reference arrows for tool
and tape scale orientation
Spring-loaded contact cylinder
Tape scale exit channel
Adhesive backing exit before roller
Applicator Tool
Tape scale entry channel
Insert screws here to attach to
fixture sensor mounting holes
Insert tape scale with adhesive backing under
the roller. Separate adhesive backing from scale
before going under roller.
Tape scale runs under applicator tool and
exits under contact cylinder.
Top View
Bottom View