13.0 Final Cleaning, Inspection, and Cure Time
Before using the encoder for servo control, clean the scale using alcohol or acetone and a lint-
free cotton cloth or swab. Finally, inspect the scale's surface for scratches, adhesive spots or
smears in the measuring length.
For tape scales, a cure time of 12 hours is required for the scale's pressure-sensitive adhesive to
achieve the best performance and reliability.
13.1 Reworking to Correct Mistakes
Once installed, the scale cannot be moved or removed and reinstalled. Reworking will require
removal and discarding of the old scale, and installation of new scale.
13.2 Cleaning Scales
Step
Action
1. General Particle Removal:
Blow off any contamination with nitrogen, clean air, or a similar gas.
2. Contamination Removal:
Use a lint-free cleanroom wipe or cotton swab dampened with isopropyl alcohol or acetone
to wipe the surface clean. Handle the scale by the edges.
Do not
scrub the scale.