MicroE Optira Installation Manual
Celera Motion
IM-1001 | REV181211
Page 10
©2018 Celera Motion
Step
Action
2. Z-Height:
If Z-Height is not adjustable, firmly place sensor against benching edge.
If Z-Height is adjustable, go to Step 5.
3. Use two M1.6 or 0-80 screws to attach the sensor and tighten per torque specification:
0.34Nm
(3 inch-lbs.) maximum.
4. Connect sensor to customer’s electronics or to the optional Connector and Calibration
Board.
Result
: Sensor installation is complete for benching install. Alignment not required with
benching installation. If sensor has an LSB index, go to
Section 5.1.5 LSB Index
Calibration.
5. If Z-Height is adjustable:
Use the Z-Height Shim Spacer (part of optional development kit) to set the proper gap
between the sensor’s riser and the top of the scale (0.493 mm). Refer to the Optira Interface
Drawing for details.
6. Insert the Z-Height Shim Spacer between the sensor’s riser and the top of the scale.
7. Fully tighten the two M1.6 or 0-80 screws to attach the sensor. Tighten screws per torque
specification: 0.34Nm (3 inch-lbs.) maximum. Adjust the Z-Height. The shim should move
with minimal friction between the riser and the scale. Remove shim.
8. Connect sensor to customer’s electronics or to the optional Connector and Calibration
Board.
Result
: Sensor installation is complete for adjustable Z-Height install.
Note
: If sensor was not installed using benching, then go to
Section 5.1.4 Sensor Head
Alignment
Sensor Head Alignment to align the sensor.
Z-Height Shim Spacer
Side View of Sensor
Riser
Installed tape scale
Side View of Sensor
Insert the shim here between
bottom of riser and top of the
installed tape scale
Installed tape scale
Riser
Z-Height Shim Spacer
Benching edge