![Norland Elite 437A150 Operator'S Manual Download Page 85](http://html1.mh-extra.com/html/norland/elite-437a150/elite-437a150_operators-manual_1705413085.webp)
Operator's Guide
437D140 Rev. I
Installation and Setup
3-29
Note:
If the
Scan Speed
is too fast for the chosen
Resolution
, the following message will appear at
the beginning of a scan:
If this occurs, the
Scan Speed
must be reduced or the
Resolution
must be increased.
Scan Width: entered in centimeters.
Minimum for
Scout
and
Measure
scans is 2 x point resolution
Maximum for
Scout
and
Measure
scans is 255 x point resolution
Note:
If the
Scan Width
is too large for the chosen
Resolution
, the following message will appear at
the beginning of a scan:
If this occurs, the
Scan Width
must be reduced or the
Resolution
must be increased.
Scout Scan: when enabled, provides image to assist in defining the Measure scan region
Force initial scan on-axis: forces end point of initial scan to be on same Y-axis as the start point
Maximum degrees to snap final scan on-axis: This field defines the maximum degree of rotation of the
cursor box at the scan progress screen (following a scout scan) that will result in an on-axis scan. On-axis
scans are faster than off-axis scans.
Rotation of the cursor box to an angle greater than or equal to the
Maximum degrees
setting will result in
scan data collection at that angle. Rotation of the cursor box to an angle less than the
Maximum degrees
set-
ting will result in an on-axis scan.
On-axis scans are scans taken with the start and end points on the same y-axis.
If the
Maximum degrees to snap final scan on-axis
is set to 15, then
Force on Axis
will correct scout view dis-
play for scout scans that are up to 15 degrees off axis.
Scan Type Name: operator-defined; default is “Research” or “Small Subject”
Summary of Contents for Elite 437A150
Page 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Page 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Page 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Page 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Page 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Page 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Page 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Page 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Page 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Page 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Page 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Page 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Page 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Page 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Page 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Page 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Page 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Page 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Page 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Page 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Page 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...