4. Position the pelvic cursor next. Move the upper control points to just above the iliac crests, between the arm
and torso. Position the bottom left pelvic cursor so that the left cursor edge passes through the femoral neck
and is close to the pelvis, and the bottom edge of the pelvic cursor is just below the pubic symphysis. Pos-
ition the bottom right cursor similarly on the opposite side. Make sure the trapezoid is set to enclose the pel-
vis.
5. If positioned correctly, the pelvic cursor will completely surround the pelvis and contain a minimum of midriff,
leg, and femoral neck tissue.
6. Position the lower left leg cursor so that the left leg cursor edge lies along the region of minimum body tissue
between the left thigh and the left hand. The bottom edge should be below the toes. Position the right leg con-
trol points similarly. Position the center leg cursor so that the center leg cursor edge lies along the region of
minimum body tissue between the left and right leg.
The regions of interest are defined graphically in Figure 5-4.
Operator's Guide
437D140 Rev. I
5-16
Scanning Whole Body
Summary of Contents for Elite 437A150
Page 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Page 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Page 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Page 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Page 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Page 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Page 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Page 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Page 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Page 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Page 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Page 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Page 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Page 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Page 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Page 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Page 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Page 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Page 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Page 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Page 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...