Operator's Guide
437D140 Rev. I
Scanning Hip
8-5
General Patient Scanning Cautions
Caution:
Properly Mark the Patient. To ensure scanner arm does not contact the patient, always
verify patient is positioned properly before scanning or moving the scanner arm.
Caution:
Do not move the patient while marking the regions to be scanned. Always remain near the
patient, in the event assistance is needed.
Caution:
Do not touch the patient and the computer system at the same time as this could increase
leakage currents.
Caution:
Do not reach around to the back of the unit while the scanner arm is moving. While guards
are provided, it is wise to avoid any chance of pinching the arm, hand, or fingers between the scanner
arm and the frame, or between the source and the scanner arm.
Caution:
Do not allow the patient to bump, push, or lean on the scanner arm. Manually moving the
arm could result in an error message which will require removing the patient from the table and doing
the Find Table Dimensions routine.
Caution:
Make certain the patient does not dangle their arm or hand over the riser while the scanner
arm is moving during a scan. The scan will not be usable, as the patient will not be properly positioned,
and the patient may be at risk of pinching their hand or finger between the scanner arm and the riser or
between the x-ray source and the scanner arm.
Caution:
Make certain the patient does not stick a finger into the slot in the bottom of the upper arm
cover during a scan; it could be pinched.
Caution:
When positioning the patient, ensure they start by sitting near the center of the table and
then swing their legs up. Sitting at either end makes positioning awkward.
Caution:
Caution the patient to remain still during the scan to ensure quality results.
Caution:
Help the patient up from the scanner after scan data collection; some patients may require a
few minutes to regain equilibrium after lying down for a length of time.
Summary of Contents for Elite 437A150
Page 1: ...DXA Bone Densitometer Operator s Guide Elite Model 437A150 Operator s Guide 437D140 Rev I...
Page 22: ...This page intentionally left blank Operator s Guide 437D140 Rev I 1 8...
Page 56: ...This page intentionally left blank Operator s Guide 437D140 Rev I 2 34...
Page 66: ...11 Click Operator s Guide 437D140 Rev I 3 10 Installation and Setup...
Page 108: ...This page intentionally left blank Operator s Guide 437D140 Rev I 3 52...
Page 124: ...Sample Printed QA Results Report Operator s Guide 437D140 Rev I 4 16 Basic Operation...
Page 137: ...Operator s Guide 437D140 Rev I Basic Operation 4 29...
Page 166: ...This page intentionally left blank Operator s Guide 437D140 Rev I 5 24...
Page 198: ...Operator s Guide 437D140 Rev I 7 20 Scanning AP Spine...
Page 206: ...This page intentionally left blank Operator s Guide 437D140 Rev I 7 28...
Page 238: ...Figure 8 6 Page 2 of the Bone Exam Report Operator s Guide 437D140 Rev I 8 32 Scanning Hip...
Page 290: ...This page intentionally left blank Operator s Guide 437D140 Rev I 10 30...
Page 326: ...Operator s Guide 437D140 Rev I 11 36 Additional Techniques...
Page 412: ...This page intentionally left blank Operator s Guide 437D140 Rev I 14 16...
Page 428: ...This page intentionally left blank Operator s Guide 437D140 Rev I 16 10...
Page 446: ...Label Locations continued Operator s Guide 437D140 Rev I 17 18 Technical Reference...
Page 448: ...Scanner Labels continued Operator s Guide 437D140 Rev I 17 20 Technical Reference...
Page 449: ...Operator s Guide 437D140 Rev I Technical Reference 17 21 Scanner Labels continued...
Page 459: ...Operator s Guide 437D140 Rev I Reference Data Sets 18 9...
Page 466: ...Operator s Guide 437D140 Rev I This page intentionally left blank...
Page 484: ...Operator s Guide 437D140 Rev I This page intentionally left blank...