2.6.2.11 Memory‑Time‑Pressure‑Vacuum Set Command (continued)
The above text packet in hexadecimal format:
STX
1
9
E
M
space space
C
H
0
0
1
0x02
0x31
0x39
0x45
0x4D
0x20
0x20
0x43
0x48
0x30
0x30
0x31
T
1
0
1
2
5
P
0
3
0
0
V
0x54
0x31
0x30
0x31
0x32
0x35
0x50
0x30
0x33
0x30
0x30
0x56
0
1
0
0
3
1
ETX
0x30
0x31
0x30
0x30
0x33
0x31
0x03
2.6.2.12 Pressure Units Set Command
This command sets the unit of measure for pressure display.
Client command and data:
E6--uu
uu:
The pressure units. 00 = PSI, 01 = BAR, 02 = KPA
EXAMPLE:
To display pressure in kPA, the text packet is: (STX) + E6-- + 02 + 7D + (ETX)
The above text packet in hexadecimal format:
STX
0
6
E
6
space space
0
2
7
D
ETX
0x02
0x30
0x36
0x45
0x36
0x20
0x20
0x30
0x32
0x37
0x44
0x03
2.6.2.13 Vacuum Units Set Command
This command sets the unit of measure for vacuum display.
Client command and data:
E7--uu
uu:
The vacuum units. 00 = KPA, 01 = Inches H
2
O, 02 = Inches Hg, 03 = mmHg, 04 = TORR
EXAMPLE:
To display vacuum in H
2
O, the text packet is: (STX) + 06 + E7-- + 01 + 7D + (ETX)
The above text packet in hexadecimal format:
STX
0
6
E
7
space space
0
1
7
D
ETX
0x02
0x30
0x36
0x45
0x37
0x20
0x20
0x30
0x31
0x37
0x44
0x03
2.6.2.14 Dispense Parameter Memory Clear
This command re-initializes the dispensing parameter memory locations by setting them all to 0.
Client command and data:
(STX) + 04 + CL-- + CD + (ETX)
The above text packet in hexadecimal format:
STX
0
4
C
L
space space
C
D
ETX
0x02
0x30
0x34
0x43
0x4C
0x20
0x20
0x43
0x44
0x03
Appendix B — RS‑232 Protocol (continued)
Ultimus V High Precision Fluid Dispenser
56
www.nordsonefd.com [email protected]
800-556-3484 Sales and service of Nordson EFD dispensing systems are available worldwide.