4-34
Power-Up
and
Testing
5.
In the FmXP Main Window, click on the
Configuration
button and select
Setup Scale
.
6.
In the Scale Setup dialog box (Figure 4-23), click on the
Scale Operations
tab.
7.
In the Scale Operations dialog box, click on the
Zero Scale
button and monitor the Current
Reading portion of the dialog box.
!
The current reading should be 0.000 mg.
Figure 4-23 Scale Setup
8.
Replace the cup and scale cover.
!
The current reading should indicate an increase in weight. If not, refer to
9.
Calibrate the scale in accordance with the procedure in