95
Connecting the optical output
Figure 62
Trigger output circuit
The high and low voltage levels are defined by the following equations:
Vhigh = Vexternal
Vlow = Vexternal x 22k / (22k + R)
Connecting the optical output
Caution:
Before connecting the optical output connector, ensure to check the
connector tip using a fiber microscope. Using the microscope, check for any de
-
formities, damage, residue or other contaminants at the optical tip of the connec
-
tor. Either clean the connector or contact NKT Photonics support if a replacement
is necessary.
The fiber and its connectors are inspected and sealed in a clean environment
before shipping. For optimum system performance, follow the guidance below:
•
Avoid removing any connector caps when the outputs are not in use and
kept in storage.
•
When connecting an output fiber, avoid:
•
Excessive strain on the fiber through stretching or twisting.
•
Fiber bend diameters < 10 cm
•
Exposing the fiber and its facets to contaminants.
•
Take good care and handle output fibers and their connectors gently to
prevent damage to the fiber or its facets.
Summary of Contents for Koheras ADJUSTIK HP
Page 1: ...Item 800 611 01 Koheras ADJUSTIK HP PRODUCT GUIDE Low Noise Single Frequency Laser System...
Page 10: ...10 E Part numbers 113...
Page 12: ...12...
Page 16: ...16...
Page 18: ...18...
Page 20: ...20...
Page 36: ...Chassis labels 36...
Page 48: ...48...
Page 56: ...Menu items 56...
Page 68: ...Enabling and disabling emission 68...
Page 84: ...CONTROL Control panel 84...
Page 86: ...86...
Page 90: ...Placing the laser on a table or shelf 90...
Page 98: ...98...
Page 102: ...102...
Page 112: ...Installing CONTROL 112...
Page 114: ...114 2...
Page 115: ...Koheras ADJUSTIK Product Description Release 1 0 04 2021 W 10456...