79
CONTROL settings
the same port for transmission as for reception (i.e. as configured under the
System
port
). Default setting is 0. If
Local port
in CONTROL’s Ethernet connection
is configured different than the
Remote
port
, the computer’s port should match
the
Remote port
.
MAC Address
The
MAC address
for the laser’s Ethernet frames is shown in the bottom of the
Ethernet settings. The
MAC address
is not configurable but unique for every
laser.
Figure 47
Ethernet setting
Watchdog
The laser features a watchdog for monitoring of USB or Ethernet communication.
If CONTROL is disconnected from the chassis, the watchdog disables emission
after its timeout expires.
Enable Watchdog
To turn the watchdog on, select ON from the drop-down menu next to
Enable
watchdog
. To turn the watchdog off, select OFF.
Watchdog timeout
When CONTROL communications is disconnected, the
Watchdog
timer starts.
When it reaches the
Watchdog timeout
setting, emission is disabled. Set
Enable
Watchdog
to ON to enable the feature.
Summary of Contents for Koheras ADJUSTIK HP
Page 1: ...Item 800 611 01 Koheras ADJUSTIK HP PRODUCT GUIDE Low Noise Single Frequency Laser System...
Page 10: ...10 E Part numbers 113...
Page 12: ...12...
Page 16: ...16...
Page 18: ...18...
Page 20: ...20...
Page 36: ...Chassis labels 36...
Page 48: ...48...
Page 56: ...Menu items 56...
Page 68: ...Enabling and disabling emission 68...
Page 84: ...CONTROL Control panel 84...
Page 86: ...86...
Page 90: ...Placing the laser on a table or shelf 90...
Page 98: ...98...
Page 102: ...102...
Page 112: ...Installing CONTROL 112...
Page 114: ...114 2...
Page 115: ...Koheras ADJUSTIK Product Description Release 1 0 04 2021 W 10456...