CONTROL overview
48
leads to an increase in side band power level and the power in the main peak
starts to decrease. This mode results in slightly higher output power.
However, power coupling into a fiber deliver system is less efficient due to
higher side band power.
Caution:
Only use
Overdrive
mode when either a new AOTF crystal or new Exter
-
nal RF Driver is added to the SELECT. In these cases, the optimum RF power cal
-
ibration curve, set at the factory, is no longer valid and a new calibration may be
required as described in
“Optimum RF power level” on page 41
. Using the
Over
-
drive
mode, results in slightly higher RF power applied to the crystal which can
cause issues with the beam lobes.
Power monitor selector
Range:
VIR/NIR
or
NIR/IR
Feedback power locking of the SuperK laser can be deployed using the
optional monitor output port of the SELECT. The port is connected to a
SuperK laser external feedback port feeding the SELECT’s power monitor
signal to the laser. The monitor port is set to supply a signal representing the
level of either the VIS/NIR or NIR/IR output port only. External power
monitoring is optional and installed at the factory – see
.
Note:
The power monitor signal can only represent the level of one of the output
ports as configured by the
Power monitor selector
setting.
VIS/NIR
or
NIR/IR power monitor gain
Range:
X1
to
X128
These settings increase or decrease the level of the monitor signal output
from the monitor output port. A gain of up to 128 times can be set and the
setting only functions when a least one power monitor and a monitor output
port is installed on the SELECT – see
.
VIS/NIR
or
NIR/IR modulation gain
Range:
0-1 V, 0-5 V, 0-10 V
When operating using
Fast wavelength switching
mode, the input
modulation gain should match modulation input signal(s) connected to an
optional SuperK COMMAND device. This setting allows you to choose
between 3 different input voltage ranges for the Modulation inputs. The
SuperK COMMAND is an optional device specifically designed for Fast
Wavelength Switching using an FSK signal. For more information refer to
Summary of Contents for A203-000-001
Page 1: ...SuperK SELECT PRODUCT GUIDE Multi Channel Tunable Filter...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 30: ...Status LEDs 30...
Page 56: ...Control panel 56...
Page 59: ...59 Figure 44 Mechanical dimensions External RF Driver...
Page 60: ...60...
Page 62: ...Support Contact Details 62 2...
Page 68: ...Installing CONTROL 68...
Page 77: ...SuperK SELECT Product Description Revision 1 00 10 2020 W 10456...