71
Signaling and CONTROL configuration
External signal
requirements
Electrical specifications for the COMMAND inputs are shown in
Table 10
COMMAND input specifications
Signaling and CONTROL configuration
For fast wavelength switching, CONTROL configuration determines the
wavelengths. Signals connected to the COMMAND module determine the
wavelength(s) selected. Refer to the figures in the following examples to see how
channel configuration relates to the input signals connected to the COMMAND
module from an application.
2 or 4-channel FWS
For switching 2 or 4 individual wavelengths, connect application signals to
module pins
Blank
and
shows a truth table demonstrating the
wavelengths selected based on the input signals levels from your application.
shows the configuration in CONTROL and how the configuration is
selected based on the signal levels at the two pins. To set up 2 or 4-channel fast
wavelength switching, follow the steps in
Table 11
2 or 4 channel FWS truth table
Procedure 7
Setting up 2 or 4 channel fast wavelength switching
1. In CONTROL, click on the SELECT device icon and set the
Operating mode
to
Fast Wavelength Switching
.
2. Select either the VIS/NIR or NIR/IR crystal.
3. Set the modulation gain in the Setup page to the 0-5 V range (see
NIR/IR modulation gain” on page 48
4. In each of the 4
Profile
s, configure channel 1 with a desired wavelength.
When only 2 wavelength switching is required, you can configure channel 1 in
shows channel 1 configured in each of the
4 profiles of a VIR/NIR crystal.
Note:
Do not set channels 2 to 7 in the four profiles, leave them OFF.
Pin
Value
Comment
FSK1 to FSK8
0-3.3 V or TTL
MOD1 to MOD8
0-1 V, 0-5 V, 0-10 V
Configured in CONTROL
BLANK
0-3.3 V or TTL
FSK1
Blank
Wavelength selected
Low
Low
Profile 1 Channel 1
High
Low
Profile 2 Channel 1
Low
High
Profile 3 Channel 1
High
High
Profile 4 Channel 1
Summary of Contents for A203-000-001
Page 1: ...SuperK SELECT PRODUCT GUIDE Multi Channel Tunable Filter...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 30: ...Status LEDs 30...
Page 56: ...Control panel 56...
Page 59: ...59 Figure 44 Mechanical dimensions External RF Driver...
Page 60: ...60...
Page 62: ...Support Contact Details 62 2...
Page 68: ...Installing CONTROL 68...
Page 77: ...SuperK SELECT Product Description Revision 1 00 10 2020 W 10456...