73
Signaling and CONTROL configuration
Procedure 8
Setting up four 8-channel profile switching
1. In CONTROL, click on the SELECT device icon and ensure the mode is set to
Fast Wavelength Switching
.
2. Select the VIS/NIR or NIR/IR crystal.
3. Set the modulation gain in the Setup page to the 0-5V range (see
NIR/IR modulation gain” on page 48
4. In each of up to 4
Profile
s, configure from 2 to 8
Channels
. Each profile can
have multiple desired wavelengths set ON.
shows 4 profiles each
with 8 channels configured for a VIR/NIR crystal.
5. Set the power level desired of the profile channels. This sets the power level
to be identical for the same channels in each profile.
6. Using a BNC cable, connect the MOD1 pin to the +3.3 V output pin on the
COMMAND. When profiles are switched with constant amplitude, the MOD
pin must be pulled high.
7. Connect the FSK1 and BLANK pins to your application.
8. The system is ready and profiles are selected based on the signals shown in
.
Figure 49
Fast wavelength switching: 4 profile setup
Profiles1 to 4 Selection
in
Blank
in
Test
Outputs
V+
CONTROL Channel/Profile
ch1 to 8 - profile 3
ch1 to 8 - profile 4
ch1 to 8/p1
AOTF profiles and their channels selected
ch1 to 8
profile 1
ch1 to 8
profile 2
ch1 to 8
profile 3
ch1 to 8
profile 4
V+
0V
V+
0V
FSK1
in
Blank
in
MOD1
in
Application
sig
nals
ch1 to 8/p1
ch1 to 8 - profile 2
ch1 to 8 - profile 1
Summary of Contents for A203-000-001
Page 1: ...SuperK SELECT PRODUCT GUIDE Multi Channel Tunable Filter...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 30: ...Status LEDs 30...
Page 56: ...Control panel 56...
Page 59: ...59 Figure 44 Mechanical dimensions External RF Driver...
Page 60: ...60...
Page 62: ...Support Contact Details 62 2...
Page 68: ...Installing CONTROL 68...
Page 77: ...SuperK SELECT Product Description Revision 1 00 10 2020 W 10456...