23
Electrical Interfaces
shows the electrical interfaces located on a side panel of the SELECT.
Figure 7
Electrical Interfaces ad optical input
Monitor output (optional)
The monitor output port outputs an electrical signal representing optical output
pulses from one of the optical outputs. Sampling from the VIS/NIR and NIR/IR
outputs is performed using optional power monitors that are fitted to the SELECT.
Note:
The port’s function is optional and only available on custom orders as the
monitors need to be fitted and calibrated with the SELECT. Contact NKT Photon
-
ics sales engineering for further information.
Accessory bus address selector
Multiple accessories can be connected on an External bus daisy chain from a
SuperK laser. Each accessory requires a unique bus address. Use this dial to set
the bus address of the SELECT.
The address selector can set the bus address with up to 16 addresses from 0x0
to 0xF (hex). To set accessory addresses in the system:
1. Ensure the laser and External RF Driver (s) are turned OFF.
2. On each accessory, set a unique address.
3. Power ON the laser and External RF Driver(s)
RF Power II LED
Indicates if RF power is received at the RFII input from an External RF Driver.
1
Monitor output
6 RF Power I LED
2 Accessory bus address selector
7 Bus through port (External bus output)
3
RF Power II LED
8 Bus input port (External bus input)
4
RF Power II input
9 Emission, Power and Interlock LEDs
i
i.
5 RF Power I input
7
1
2
8
3
4
5 6
9
Summary of Contents for A203-000-001
Page 1: ...SuperK SELECT PRODUCT GUIDE Multi Channel Tunable Filter...
Page 8: ...8...
Page 10: ...10...
Page 14: ...14...
Page 30: ...Status LEDs 30...
Page 56: ...Control panel 56...
Page 59: ...59 Figure 44 Mechanical dimensions External RF Driver...
Page 60: ...60...
Page 62: ...Support Contact Details 62 2...
Page 68: ...Installing CONTROL 68...
Page 77: ...SuperK SELECT Product Description Revision 1 00 10 2020 W 10456...