Instruction Manual MA-3000+RD-5+SC-5
NIC-TD-0000015-05
45
5.5. Trouble shooting
Bubbling is weak
?
Reducing agent is new?
Replace reagent pump or
tubes
Peak sensitivity
is recovered?
Contact our
service
department
START
END
YES
YES
Readjust standard solution
Standard solution
is correct?
Exchange new tube
Wash tubes by alkaline
cleaning
Clean up optical cell
Readjust reducing agent
YES
YES
YES
YES
NO
NO
NO
NO
NO
NO
Check tube connection
Test tubes are clean
?
NO
YES
YES
No bubbles?
Replace mist catcher
NO
Peak shapes are normal?
The reagent is dropped
?