Instruction Manual MA-3000+RD-5+SC-5
NIC-TD-0000015-05
41
5.4.3. Cap unit tubes and seal sensor
1) Carry out the reagent collection using the MA-3000.
2) Turn off the main detector and sample changer.
3) Loosen the two resin screws and remove the cap unit cover.
4) When replacing the seal sensor, remove the two resin screws fixing the
sensor, and remove it from the arm block.
Loosen the three resin screws and separate the arm block and the cap unit.
5) Loosen the three resin screws of the cap unit.
Cap unit cover
Resin screw
(also on opposite side)
Cap unit
Resin screw (3 places)
Seal sensor
Resin screw for fixing sensor
(2 places)
Arm block
Resin screw
(3 places)