Instruction Manual MA-3000+RD-5+SC-5
NIC-TD-0000015-05
34
4.10. Termination of measurement
(1)
To wash reagent tubes. Select MENU “Run (R)” - "Reagent operation" - "Reagent
discharge"
*
Please follow to the message window
on PC.
(2)
Discard the sample in WASTE bottle and waste fluid tank.
(3)
Remove and clean a turntable.
(4)
The “Setup” window from “System”-“Setup” is opened and return measuring method
to “THERMAL DECOMPOSITION”.
(5)
Turn off MA-3 and SC-5.
(6)
End software.
4.11. Washing the test tube and bubbler
If SnCl
2
adheres in the tube and bubbler, mercury may be lost before sample analyzes.
The used test tubes and bubbler are required to wash by the Alkali detergent or
0.1M-NaOH for 30 minutes using Ultrasonic cleaner and wash by (1+10) Nitric Acid for
30 minutes using Ultrasonic cleaner.