Instruction Manual MA-3000+RD-5+SC-5
NIC-TD-0000015-05
47
6.2. SC-5
6.2.1. General performance
Number of sample
containers
Refer to the specification of the turntable in the later part.
Sample access
Random access
Bubbler tube
PEEK tube
Glass bubbler in the case of the optional large capacity
Washing of bubbler
tube
Overflow washing in washing tank using pure water
Sample container
detection
Presence detection using sensor
Control method
Communication control using RS-232C
6.2.2. Performance of each part
6.2.2.1. Turntable
Number of sample
containers
Standard table: 30 dedicated sample containers for 5 mL
Driving method
Stepping motor
Reference position
Table No. 1
Stop position
Each sample container position
6.2.2.2. Arm rotation
Driving method
Stepping motor
Reference position
Washing tank
Stop position
Washing tank and measurement position
6.2.2.3. Vertical movement of arm
Driving method
Stepping motor
Reference position
"UP" position
Stop position
"UP" position, measurement position, washing position,
head-up position, measurement preparation position